SOP for monitoring temparaturue of degital thermometer of BOD Incubator and Thermostatic Incubators.
OBJECTIVE :- To establish a procedure for temperature monitoring of Digital
Temperature BOD Incubator and Thermostatic Incubators
UNIT OF RESPONSIBILITY :- QA Executive and Trained QA Assistants
PROCEDURE :-
1) See that the incubators are in ON position
2) Check the temperature to ascertain that it is within the specified limits
3) Check the temperature shown by digital temperature indicator of BOD and
thermostatic incubators
4) The temperature shown by the incubators should not differ by + 05 deg C from the
set temperature
5) Maintain the records of temperatures as per Attachment-I in respective QA,File
TOLERANCE LIMITS : – + 05 deg C
OUT OF LIMITS ACTION :-
If the temperature is not within the specified limits, adjust with the help of adjustment knob
MAINTAINENCE /SERVICING :- To be done as and when required
PERSONNEL TO BE REPORTED TO IN CASE OF NON CONFORMANCE /DEVIATION :
Maintenance Department
FREQUENCY OF OPERATION : Thrice a day
NO OF ATTACHMENT :
RETENTION PERIOD OF RECORDS : Five years
This Website is a Guide for Pharmaceutical Manufacturing Pharmacy Students Pharmacy Colleges and Pharmacists pharmaceutical companies in health care